Using the Widener University VG-70 SE
Overview of instrument
General Diagram:
Source Region:
Collision Cell:
Electrostatic Analyzer:
Detector:
Power on computer, electronics, and heaters
- Electronics on
- Power on to electronics cabinet (rockerswitch on back of instrument)
- Power On (switch on front of console, shown below)
- Set the magnet energy to IB
- Heaters and ion gauges
- Set inlet reservoir and transfer line heaters.
- S REEN Transfer line feedthrough °200 C
- S RES Inlet Reservoir 200 °C (max temp achieved)
- Set the source temp, 200 °C for typical operation, 300 C max, gauge reads higher
than setpoint
- Turn on Ion gauges to "Read" (switch analyzer gauge first, shown below)
Computer on
- turn on power strip (behind GC)
- LOGIN VG
- PASSWORD analysis
- to load opus software from menu select: applications; decterm
- NOTE: When entering a value in a window, always use the enter key so that
value is accepted.
Operate
- Electron Ionization (controls shown below)
- Operate, turns on electron filament & high voltage
- Set the electron beam energy
- The emission current should be ca 2 mA
- Set the electron multiplier, 3 to 4 kV
- In opus window, click on instrument icon to see settings for
- acceleration voltage (8 kV)
- electric sector voltage (8 kV)
- magnet (m/z to transmit, 69 is good starting point)
- Observe signal with scope (controls shown below)
- use span and mv/div to observe peak on scope
- set filter to 0.3
- use keyboard to enter approximate mass in box for the magnet in the opus
instrument window
- use L & R arrow keys to shift settings so peak is on scope
- calculate resolution as: 106/span (in ppm at 5% peak height from scope)
To load sample. (1 ul will last ca 1 hr)
- For solids probe Tmax = 350 °C
- Direct inlet reservoir
- The sample reservoir has 2 valves (see diagram below)
- Pump out reservoir open both valves (Right valve first)
- Run sample close right valve
- Open reservoir close both valves
- Valves are on/off, no control or needle
- Right side valve goes to the pump
- Left side valve goes to chamber
Tune instrument
- Tune Source to maximize signal.
adjust corse, fine, & deflect.
controls are interactive.
- 1 is source
- 2 is collision cell
- Adjustments
- analyzer
- y, z, z1, z2, zfocus and deflect
- before magnet
- Curve 1
- Rotate 1
- after magnet
- Curve 2
- Rotate 2
- change slits as needed
- Close CCW to turn up
- Open CW to turn down
- Tune at low resolution
- Open up beam restrictors
- Open the source slit
- Open up collector slit so peak has flat top
- Open alpha slit, and z restrictors
- Tune for maximum intensity
- Close collector to obtain peak shape at ca 70% intensity and tune
- For high resolution
- Open collector to flat top
- Close source to ca 10% intensity (3 clicks in mV) and tune
- Close collector to ^ peak and tune curve, y-focus & rotation
Calibration
- Experiment
- Set Limits
- For exact mass scan very slowly, suggest Time = 20 s for scan from m/z 10 to 300.
- For exact mass set to continuum mode.
- Blank out current calibration (>> Button)
- Inject 1 uL PFK PFK Spectrum or Table
- Acquire data
- File: Cal File Name
- Data: Location Data
- Overwrite
- Load
- Start
- Calibration window comes up
- REF, directory, select PFK
- Get a good PFK Scan for calibration reference, need good S/N
- Majority of peaks should be close
- Major peaks are shown in spectrum below
- Adjust low and High values so peaks match close to expected values
- Change
- Return
- Automatic
- Adjust location and predict values to increase number of peaks found.
- When adjusted Overwrite; Accept
Acquiring High Resolution Data
- Experiment settings. Same as setup for calibration, any changes require recalibration.
- Set Limits
- For exact mass scan very slowly, suggest Time = 20 s
- For exact mass set to continuum mode.
- Acquire
- File Name
- Text Box, add description of experiment
- Overwrite
- Load
- Start
- Acquire 10 scans with just PFK
- Inject sample
- For realtime data
- File; open, select file
- chromatagram (total ion current)
- click on clock for reatime data
- click on chromatagram for mass spectrum
- To stop acquisition, acquire; stop
- Bring up data with Histogram button
Processing High Resolution Data
- File: Open
- Click on Chromatogram
- Note for mouse movement
- Middle Drag to average sample + PFK
- Left drag to average background (PFK)
- Right click to subtract and display
- On Spectrum
- Annotate
- Modify
- Decimal Places (set to 4)
- To set calibration
- In chromatagram, middle drag on PFK and right click to send to spectrum
- Histogram data, Spectrum; utilities; peak detect
- Utility; Calibrate
- Change calibration file name (use filename_1)
- Auto calibrate
- overwrite
- accept
- In opus window, Analysis; Mass Measure
- Select Reference Compound (PFK)
- Select calibration file named above
- Click Accurate
- Set Windows
- Initial ppm 400
- Final PPM 200
- To analyze sample, middle drag on sample, left drag on PFK, right click to display spectrum
- Histogram data, Spectrum; utilities; peak detect (for close peaks set minimum width to 1 in peak detect options)
- Number peaks, Annotate; modify.
- zoom in on peak of interest
- For elemental analysis
- In opus window, Analysis; elemental
- Control; elements (select elements to search and specify allowed range and
isotopic abundance)
- Control; show (to select minimum and maximum abundance for peaks to analyze)
- graph; display range to calculate elemental composition of peaks in spectrum.
- Printing does not currently work from this window.
- Exact mass calculator from SIS
- Exact mass of elements from SIS
- PFK reference spectrum from SIS
- If you receive a data error message, close and reopen OPUS
To power down Instrument
- Set to electronics to Standby
- Source temp to 0 C
- Inlet & Transfer line to 75 C
- Ion Gauges OFF
- Pump out sample reservoir (open right valve)
- Power off
- To Shutdown Computer
- Close opus software
- from session manager window, menu select: session; end session
- login as
- LOGIN system
- PASSWORD security
- to shutdown select: application; decterm; 13
- AT <<< Prompt turn off power strip to computer
- Power off to electronic console
This page is maintained by
Scott Van Bramer
Department of Chemistry
Widener University
Chester, PA 19013
Please send any comments, corrections, or suggestions to
svanbram@science.widener.edu.
This page has been accessed
1788
times since 5/30/97.
Last Updated Wednesday, October 11, 2000 21:27:33