VACUUM SYSTEM:
All mass spectrometers operate at very low pressure (high vacuum). This reduces the chance of ions colliding with other molecules in the mass analyzer. Any collision can cause the ions to react, neutralize, scatter, or fragment. All these processes will interfere with the mass spectrum. To minimize collisions, experiments are conducted under high vacuum conditions, typically 10-2 to 10-5 Pa (10-4 to 10-7 torr) depending upon the geometry of the instrument. This high vacuum requires two pumping stages. The first stage is a mechanical pump that provides rough vacuum down to 0.1 Pa (10-3 torr). The second stage uses diffusion pumps or turbomolecular pumps to provide high vacuum. ICR instruments have even higher vacuum requirements and often include a cryogenic pump for a third pumping stage. The pumping system is an important part of any mass spectrometer but a detailed discussion is beyond the scope of this paper.